DICOM is a registered trademark of the National Electrical Manufacturers
Association.
The system conforms to the DICOM standard as specified in the Vevo MD Imaging
System DICOM Conformance
Statement.
This statement provides information about
the purpose, characteristics, configuration, and specifications of the network
connections supported by the system. For more information contact FUJIFILM
SonoSite.
HIPAA standard
The system includes security settings that help you to meet the applicable security
requirements listed in the HIPAA standard. Users are ultimately responsible for
ensuring the security and protection of all electronic protected health information
collected, stored, reviewed, and transmitted on the system.
The Health Insurance and Portability and Accountability Act, Pub.L. No. 104-191
(1996). 45 CFR 160, General Administrative Requirements.
45 CFR 164, Security and Privacy.
Measurement accuracy
The measurements from the system are of a physical property such as distance for
evaluation by the clinician. The accuracy values require that you can place the
calipers over one pixel. The values do not include acoustic anomalies of the body.
The 2D linear distance measurement results are displayed in millimeters with one
place past the decimal point.
The linear distance measurement components have the accuracy and range shown in
the following tables.
2D Measurement Accuracy and Range
2D
Measurement
System
Accuracy By
Test Method
Range
a
Accuracy and
Tolerance
Range
Linear distance ˂± 2% plus 1% Acquisition
Phantom
0-37.7 mm
HIPAA standard
407