Rated maximum
output power
of transmitter
(Watts)
1
10
100
100
For transmitters rated at a maximum output power not listed above, the
recommended separation distance (d) in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note:
At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies. These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects, and
people.
Separation distance (continued)
Separation distance according to frequency of
transmitter (m)
150 kHz to 80
80 MHz to 800
MHz
d=1.2 √P
1.2
3.8
12
12
800 MHz to 2.5
MHz
d=1.2 √P
1.2
3.8
12
12
Electromagnetic compatibility
GHz
d=2.3 √P
2.3
7.3
23
23