FujiFilm VisualSonics Vevo MD User Manual page 353

Hide thumbs Also See for VisualSonics Vevo MD:
Table of Contents

Advertisement

WARNING:
The power cord that connects the ultrasound system to mains
power must only be used with the system, and cannot be used to connect
other devices to mains power to avoid the risk of electrical shock and fire
hazard.
WARNING:
To prevent injury to the operator/bystander, the transducer must
be removed from patient contact before the application of a high-voltage
defibrillation pulse.
WARNING:
To avoid possible electrical shock or electromagnetic
interference, verify proper operation and compliance with relevant safety
standards for all equipment before clinical use. Connecting additional
equipment to the ultrasound system constitutes configuring a medical
system. FUJIFILM SonoSite recommends verifying that the system, all
combinations of equipment, and accessories connected to the ultrasound
system comply with JACHO installation requirements and/or safety
standards such as AAMI-ES1 60601-1, NFPA 99 or EN/IEC Standard 60601-1
and electromagnetic compatibility standard IEC 60601-1-2 (Electromagnetic
compatibility), and are certified according to IEC Standard 60950
Information Technology Equipment (ITE) or IEC 60601-1.
WARNING:
Because the only method of completely removing AC power
from the Vevo MD Imaging System is to disconnect the AC input power
cord from the stand base, ensure that you place the system in a location in
which you can easily remove the AC input power cord if necessary.
WARNING:
Failures in the electrical safety design of connected devices
may result in a voltage on the ultrasound system. To minimize the risk of
electrical shock to the patient and/or operator: Use medical-grade devices;
After connections are made, test the electrical safety utilizing biomedical
department electrical safety procedures; For added protection, use an
isolation transformer between the connecting device and AC power.
Caution:
Do not use the system if an error message appears on the clinical display.
Note the error code and call FUJIFILM SonoSite or your local representative, turn
off the system by pressing the power button until the system powers down.
Electrical safety
353