b. If you are in completed exams, select an exam and tap Review Images, then
double tap a thumbnail image to open in review—If the image is a clip, pause
the clip
c. If you are acquiring image data, tap Freeze on the control panel
2 View the available measurements by tapping the Measurements button. Only the
available measurements for the active mode will be available.
WARNING:
To avoid misdiagnosis or harming the patient outcome, start a
new patient form before starting a new patient exam and performing
measurement calculations. Starting a new patient form clears the previous
patient's data. See
WARNING:
To avoid incorrect measurement calculations, verify that the
patient information, and system date and time settings are accurate and
correct.
WARNING:
To avoid misdiagnosis or harming the patient outcome, do not
use single calculations as sole diagnostic criteria. Use calculations in
conjunction with other clinical information.
WARNING:
The Vevo MD Imaging System uses ultra high frequency (UHF)
series transducers. Each UHF transducer model has a different image scale
that does not start at 0.0 mm, but includes an offset. Please be aware of this
when placing measurements.
To place a measurement on an image
1 Tap Measurements.
284
Measurements
"Entering patient information manually"
on page 173.