Troubleshooting - Siare Morpheus Series User Manual

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

The losses value of anaesthesia unit is displayed in ml basing on known
and calculated data (flow, pressure and time),
The displayed value must be less than 100 ml.
The values obtained during the LEAK TEST are displayed for the
"compensation of dead space" function.
Leakage TEST not overcome
WARNING – Risk of failure of anaesthesia unit.
This test is not overcome when the pneumatic circuit pressure does not
reach 30 cmH 2 O ( code
code
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
E2
).
If the preliminary checks phase is not overcome: see cfr. 8 (Alarms)
and cfr. 9 (Troubleshooting).
E1
) or when there is a leakage over 100 ml (
AD.A module
5-11