3.6.6 Preliminary checks – Alarms
3-28
Ventilator operation check
Make sure that the airways pressure increases during the inspiratory
phase.
Make sure that the airways limit pressure intervenes (pressometric
operating mode).
Make sure that the variation in the set oxygen concentration value (O 2
%) corresponds.
Make sure that the ventilator responds properly at parameters variation.
Make sure that that the trigger works properly.
Make sure that the values set for respiratory frequency and volume are
properly displayed and the pressure, volume and flow curves match the
monitored parameters.
Make sure that the alarms intervene properly.
If the O
measured value differs from the set value by more than +/-
2
10%, please repeat the "O
DEMAND.
• If the Vte measured value differs from the set value by more than +/-
20% ( ADULTS parameters ), please go to the "CALIBRATION
PROGRAMS" visualization to perform the calibration of the expiratory
flow sensor (see chapter 3.8 and/or 4.13).
To carry out the preliminary checks relative to ventilator's alarms, you need
to know how the keyboard and the alarms work: please refer to the relative
chapters.
Sensor Calibration" procedure: TEST ON