Preparation to use
6.5.4 Preliminary tests - Alarms
Alarms Test
•
Low pressure
•
High pressure
•
High/low oxygen
concentration
6-48
To perform the preliminary tests on lung ventilator alarms it is
necessary to have knowledge of the keypad and alarm logic
operation (see cf. 5 and/or cfr. 8).
For additional information on alarm operation logic consult cfr. 8.
•
During anaesthesia unit operation discounnect the
patient simulator from patient circuit (SIARE type).
•
After around 15 seconds the low airways pressure
alarm occurs: silence the alarm.
•
Re-connect the patient simulator.
•
Set the high pressure alarm limit at 45 cmH 2 O.
•
During the anaesthesia unit operation block (wit the
hands) the patient simulator.
•
The high airways pressure alarm occurs: silence the
alarm.
•
Unlock the patient simulator.
See cfr. 6.5.1