Preparation to use

6.5.4 Preliminary tests - Alarms

Alarms Test
Low pressure
High pressure
High/low oxygen
concentration
6-48
To perform the preliminary tests on lung ventilator alarms it is
necessary to have knowledge of the keypad and alarm logic
operation (see cf. 5 and/or cfr. 8).
For additional information on alarm operation logic consult cfr. 8.
During anaesthesia unit operation discounnect the
patient simulator from patient circuit (SIARE type).
After around 15 seconds the low airways pressure
alarm occurs: silence the alarm.
Re-connect the patient simulator.
Set the high pressure alarm limit at 45 cmH 2 O.
During the anaesthesia unit operation block (wit the
hands) the patient simulator.
The high airways pressure alarm occurs: silence the
alarm.
Unlock the patient simulator.
See cfr. 6.5.1