Table of Contents

Advertisement

3.3

Use

To obtain better performances, leave the lung ventilator working for at least
15 minutes, before patient connection or before executing preliminary
checks.
This operation will allow the system to reach the correct temperature for the
spirometry.

3.3.1 Preliminary tests

Before using the lung ventilator on a patient, it is necessary to perform some of preliminary
tests in order to check that the equipment is properly operating.
The preliminary checks have the aim to verify the correct connection and functionality of the
lung ventilator and all its components.
The list of preliminary tests is available at the end of the present chapter or in
APPENDIX chapter.
The preliminary tests should be performed:
Before starting preliminary tests, the Falco 202 Evo lung ventilator must be:
prepared for use (cfr. Maintenance, Cleaning, Disinfection and Sterilisation)
correctly positioned
all accessories and devices for correct operation must be predisposed
electric power and gas supplies must be connected
it is used a patient simulator fixed to the terminal of supplied patient circuit.
The patient simulator suggested for tests and checking's is SIARE code
LS.AB.001 which is equipped with variable compliance and resistance.
3-14
each time the lung ventilator is turned ON and used
or whenever a connection is made or an important component is
replaced (patient circuit, oxygen probe, flow sensor, etc...)
User manual, DU3104101