Mains Power Supply - Siare Morpheus Series User Manual

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

Preparation to use

6.3.7 Mains power supply

The electrical connections are a very important part in the installation of the
anaesthesia unit. Incorrect connections or connections to unsuitable electrical
systems can compromise the safety of the patient and the operator.
The mains power supply must comply with the CEI 64-8/7 regulations concerning
type A premises for medical use.
The power supplies foreseen on MORPHEUS LT/MRI anaesthesia unit are of two
types:
by main power supply;
by internal battery.
Connection to the mains power supply
The main electric power supply must
correspond to that indicated on the
identification label (electric power supply,
frequency and power consumption) located
on the back side of anaesthesia unit:
100 ÷ 240Vac / 47 ÷ 63Hz / 200W
connect the power cable outlet to the
plug on the anaesthesia unit.
insert the plug of the power cable in the
main power supply outlet.
position the main switch (located in the
back side of the anaesthesia unit) in
position "I".