4.3.6 Patient circuit connections

Insert the patient circuit supporting arm in the specific bracket, fixed to the full-height steel rod used
for accessories mounting.
CAUTION
Use a suitable patient circuit for the
patient ventilation.
Connect the patient circuit to the specific INS.
and EXP. connectors on the valves group.
Position the patient circuit on the patient circuit
supporting arm.
WARNING !! Risk for Patient injury.
WARNING !! Risk of strangling
Pay special warning when connecting the patient circuit to the anaesthesia unit. If
not carefully positioned, the hoses, the cables, the patient circuit and other similar
components to the anaesthesia unit, these can be dangerous for the patient.
WARNING !! Risk of burns
Do not use conductive masks or conductive breathing hoses during surgery with
electrosurgical units because they can cause burns.
4-10
Whenever the circuit is changed or replaced, it is necessary to perform the Leak
Test to verify eventual leaks and the compliance of patient circuit ( please see
chapter. 4.8 ).
The system checks the patient circuit at every start-up of the unit: Self Test
phase, see on chapter 5.2 or chapter 4.7 )
Tidal Volume
< 50 mL
from 50 to 200 mL
> 210 mL
Set of hoses
Neonatal
Paediatrics
Adults