6.2.4 Battery charger

The anaesthesia unit is equipped with an internal battery (Pb 12Vdc 3Ah) that
guarantees (if perfectly efficient) at least 120 minutes operation (90 minutes,
according to ventilation parameters), in case of power failure.
The level of battery charge is constantly controlled so, in case the residual
battery is low, the switching to battery operation is made automatically: At the
same time, the relevant alarm is activated and on the ventilator screen appears
the relevant message: "Power failure".
The internal battery can be recharged by connecting the anaesthesia unit to
main power supply (using power cable supplied with the unit).
Carry a battery charge at least 10 hours, before using the the
anaesthesia unit the first time.
The operating time of anaesthesia unit guaranteed by the battery, can
vary in the following cases:
User's Manual, DU3310103-MRI / rev.3 - 02.05.2019
old battery or not perfectly efficient;
not standard ventilatory parameters;
presence of electronic flowmeter box.