4.2.3 Battery charger

The anaesthesia unit is equipped with an internal battery that guarantees (if perfectly efficient) at
least 120 minutes operation (90 minutes, according to ventilatory parameters), in case of power
failure. The switching to battery operation is made automatically: on the unit screen appears the
relevant message " POWER SUPPLY FAULT ".
The anaesthesia unit battery can be recharged by connecting the equipment to main power supply
and placing the main switch in " I " position.
Carry a battery charge at least 10 hours, before using the the anaesthesia unit the
first time.
The operating time of anaesthesia unit guaranteed by the battery, can vary in the
following cases: old battery or not perfectly efficient, not standard ventilatory
parameters, presence of electronic flowmeter box.
Insert the plug of the power supply cable to the wall
socket.
Connect the power supply cable to the anaesthesia
unit.
The electric main power supply must correspond to
the one indicated in the identification label on the
back side of the unit.
Set the main switch (placed on the back side of the
anaesthesia unit) to " I ".
Verify on the front module of unit (user's control area) if the green led is light on (it indicates
the main power supply).
Morpheus_ND
To ensure maximum autonomy of operation, it is necessary to guarantee
sufficient recharging time.
To bring the charge level from 0 to 90% takes approximately 10 hours
recharging through the mains power supply.
It is not necessary to start-up the anaesthesia unit.
4-5