AD.A module
5.3
Flow sensor calibration
WARNING – Risk of failure to the anaesthesia unit and/or injuries to
the patient.
The flow sensor calibraition should be performed:
•
•
WARNING – Risk of failure of the anaesthesia unit and/or injuries to
the patient.
•
•
Flow sensor calibration
•
STAND-BY operative mode.
•
Flow sensor correctly mounted (EXP
line) and connected to the anaesthesia
unit.
•
Patient circuit connected to the valves
group with Y connector plugged.
5-20
Exclusively by qualified and SIARE trained personnel.
When there are differences higher than 20% between set volume
(Vt) and measured value of expired tidal volume (Vte).
Before performing the calibration of flow sensor execute the leak
TEST (see cfr. 5 page 10).
If the leak TEST is overcome proceed with flow sensor calibration.
MORPHEUS MRI