Flow Sensor Calibration - Siare Morpheus Series User Manual

Anesthesia unit
Hide thumbs Also See for Morpheus Series:
Table of Contents

Advertisement

AD.A module
5.3

Flow sensor calibration

WARNING – Risk of failure to the anaesthesia unit and/or injuries to
the patient.
The flow sensor calibraition should be performed:
WARNING – Risk of failure of the anaesthesia unit and/or injuries to
the patient.
Flow sensor calibration
STAND-BY operative mode.
Flow sensor correctly mounted (EXP
line) and connected to the anaesthesia
unit.
Patient circuit connected to the valves
group with Y connector plugged.
5-20
Exclusively by qualified and SIARE trained personnel.
When there are differences higher than 20% between set volume
(Vt) and measured value of expired tidal volume (Vte).
Before performing the calibration of flow sensor execute the leak
TEST (see cfr. 5 page 10).
If the leak TEST is overcome proceed with flow sensor calibration.