4.16 Calibration Programs

4.16.1 Preliminary

WARNING !! Patient / User injury hazard
WARNING !! Patient / User injury hazard
All figures and examples featured in this chapter are purely informative and
do not refer to real clinical cases.
CAUTION
Falco 202 Evo
The information's herein are exclusively intended for use by SIARE
specialised or qualified technical staff, formally authorised by SIARE.
In order to start the Calibration Program is necessary the intervention of
SIARE personnel or qualified technical personnel authorized by SIARE.
The procedures described are critical operations and must be carried out
only by authorised staff as they might affect the equipment's safety and
proper operation.
The specialist SIARE staff or qualified technical staff, formally authorized
by SIARE, must know the full content of this manual (and of the Service
manual), before carrying out the operations described below.
The SIARE authorized technician avails of suitable tools and spare parts
and is trained to work in compliance with product safety.
SIARE declines all liability for technical interventions carried out on the
equipment without formal authorisation from SIARE.
In order to start the Calibration Program, display the lung ventilator
should be operating and correctly connected.
For tests and checks, please use the patient simulator SIARE cod.
LS.AB.001 that is equipped with variable resistance and compliance.
For a more correct and detailed analysis of the Calibration Programs,
please consult the SERVICE manual.
4-69