Table 18 Display En15317 Specifications; Table 19 Transmitter En15317 Specifications; Table 20 Receiver En15317 Specifications; Table 21 Other En15317 Specifications - Olympus 45MG User Manual

Ultrasonic thickness gage
Hide thumbs Also See for 45MG:
Table of Contents

Advertisement

DMTA-10022-01EN, Rev. C, January 2015
Parameter
Type
Size
Parameter
Transmitter pulse
Pulser voltage
Pulse rise time
Pulse duration
Parameter
Gain control
Frequency range
Parameter
Data storage
Data output types
Calibration setting storage
210
Appendix A

Table 18 Display EN15317 specifications

Color graphical TFT, LCD, 320 × 240 pixels
[Height] × [Width], [Diagonal]
41.15 mm × 54.61 mm, 68.58 mm
(1.62 in. × 2.15 in., 2.70 in.)

Table 19 Transmitter EN15317 specifications

Adjustable square wave pulser
Pulse voltages: 60 V, 110 V, 150 V, and 200 V
Damping in: 5 ns typical
Damping out: 3.5 ns typical (pulse-width dependent)
Adjustable to probe frequency

Table 20 Receiver EN15317 specifications

Automatic or manual: 40 dB to 99 dB
0.5 MHz to 24 MHz typical (filter dependent)

Table 21 Other EN15317 specifications

Internal and external microSD memory cards, maximum
capacity of 2 GB.
Each card:
475000 thickness readings or
20000 waveforms with thickness readings
2.0 USB client
Removable microSD memory card
Default single and dual element transducer setups
35 custom setups (single element)
Value
Value
Value
Value

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents