NOTE
You can also turn off the OmniScan MX2 by pressing and holding the Power key for
ten seconds. However, your setup will NOT be saved.
2.2
Automatic Start-Up Mode
The OmniScan MX2 has an automatic start-up mode: auto-boot. Use the auto-boot
mode to remotely start the OmniScan MX2 instrument. When this mode is enabled,
you do not need to press the Power key (
OmniScan MX2 starts up automatically when connected to a DC power adaptor. This
mode is disabled by default.
To activate the auto-boot
1.
Turn off the OmniScan MX2 instrument, remove the batteries, and then
disconnect the DC power adaptor.
2.
Press and hold the Power key (
3.
Connect the OmniScan MX2 to a DC power adaptor.
4.
Wait for the power indicator light to blink two times, and then release the Power
key (
).
5.
To deactivate the auto-boot, repeat steps 1 to 4.
2.3
Power Supply Management
The OmniScan MX2 is a portable instrument that can draw power from either
lithium-ion batteries or a DC power adaptor.
DMTA-20015-01EN [U8778402], Rev. M, February 2017
) to start the OmniScan MX2. The
).
Basic Operation
37