DMTA-10055-01EN, Rev. A, February 2015
8.
Adjust the value to match the known thickness of the gated indication (0.200 in. in
this example), and then press Continue (P3 key) to continue to the second
calibration step (see Figure 9-3 on page 149). The thickness value used in this part
of the calibration will be stored for reference in the Cal-Zero parameter box.
To exit without acquiring calibration data, press Cancel (P1 key).
148
Chapter 9
Figure 9-2 Entering the Zero Cal thickness value