| Handling and Control
Chapter 7
|
7-2
Model 3962

Keypad Operation

The Model 3962 Series environmental chamber has four basic modes that allow
environmental chamber setup: Run, Setpoints, Calibration and System Configuration.
Run is the default mode during normal operation.
Set is used to enter system setpoints.
Calibration is used to calibrate various system parameters.
Configuration allows for custom setup of various options.

The chart below shows the selections under each of the modes.

Figure 7-2
Modes

Set the Operating Temperature

This environmental chamber has an operating temperature setpoint range of 5.0 °C
above ambient to 60.0 °C. It is shipped from the factory with a temperature setpoint of
10.0 °C. At this setting, all heaters are turned off.To change the operating temperature
setpoint:
1. Press the Mode key until the Set indicator lights.
2. Press the right arrow until "TEMP XX.XC" is displayed in the message center.
3. Press up/down until the desired temperature setpoint is displayed.
4. Press Enter to save the setpoint.
5. Press the Mode key until the Run indicator lights for Run mode or press the right/left
arrow keys to go to next/previous parameter.