Setting Shaking Speed; Calibrating Shaking Speed; Setting Operating Temperature - Thermo Scientific MaxQ Mini 4450 Manual

Hide thumbs Also See for MaxQ Mini 4450:
Table of Contents

Advertisement

Note
Speed can be changed without press-
ing the START or STOP membrane
switches. Simply press the appropri-
ate UP or DOWN switch until desired
RPM is reached.

Setting Shaking Speed

1.
Hold down appropriate arrow membrane switch
in the speed module of the control panel, up or
down, until desired speed is set up to 500 rpm.
SET RPM light will illuminate.
2.
Press START membrane switch to begin shak-
ing. RPM light will illuminate.
3.
Press STOP membrane switch to end shaking.
SET RPM light will illuminate.

Calibrating Shaking Speed

1.
Choose a speed for which calibration is desired
by using the shaker's up or down arrow keys.
2.
Measure current shaker speed by using a digital
hand held tachometer.
3.
If the tachometer reading matches the shaker
display, no calibration is necessary. If the
tachometer reading is different from the shak-
er's display, then calibration is required.
4.
To get into the calibration mode, hold down the
START key, press and release the STOP key,
then release the START key.
5.
The decimal point on the SPEED display will
flash indicating you are in the calibration mode.
6.
Use the up or down arrow keys to set the shak-
er speed to match the tachometer's readout.
7.
Press STOP key to enter the new speed value.
8.
Press START key to exit the calibration mode.

Setting Operating Temperature

1.
Press and hold up arrow key to increase tem-
perature, release key when desired set point is
obtained.
O
PERATION
19

Advertisement

Table of Contents
loading

Table of Contents