Memory Sanitization and Removal Procedures
Archiving and Restoring Factory Calibration Data Files
5. Select the calibration data file, then copy it to the Windows clipboard by selecting
6. Navigate to the folder below and paste the file by selecting Edit > Paste from the
7. Verify that the file
8. Cycle the instrument power.

Software Revision Greater than A.01.55 and Lower than A.02.00

If the instrument software revision is greater than A.01.55, the calibration data
is stored in an SQL database file.
For software revisions greater than A.01.55, but lower than A.02.00, the SQL
file is written and read by a utility program called BackupAndRestore.exe.

Data Backup using Utility Program

Follow the steps below:
1. Close the instrument application software by pressing File > Exit > Enter.
2. Plug the mouse's USB cable into one of the instrument's USB ports.
3. Plug the USB Memory Device into another of the instrument's USB ports. For details of
4. Plug the USB keyboard into another of the instrument's USB ports.
5. Using the mouse double-click on the Windows Explorer icon on the desktop and
6. Double-click on BackupAndRestore.exe.
Figure 5-4
Calibration Data Backup
Security Features & Document of Volatility
CurrentDataSet.xml
Edit > Copy from the Windows Explorer pull-down menu.
Windows Explorer pull-down menu:
E:\AlignDataStorage
CurrentDataSet.xml
appropriate devices, see
"Selection of USB Memory Device" on page
navigate to the following folder:
C:\Program Files\Agilent\SignalAnalysis\Physics
The
Calibration Data Backup And Restore
appears.
was copied to the new location.
dialog shown in
42.
Figure 5-4
45