Chapter A - GENERAL & TECHNICAL DATA
6 EMC INFORMATION
WARNING
Use of any accessories and cables other than those specified in Planmeca
Compact i dental unit's documentation, with exception of cables sold by Plan-
meca as replacement parts for internal components, may result in increased
emission or decreased immunity of the dental unit.
WARNING
Planmeca Compact i dental unit should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the Planmeca Com-
pact i dental unit should be observed to verify normal operation in configura-
tion which it will be used.
Planmeca Compact i dental unit is intended for use in the electromagnetic environment specified
below. The customer or the user of the Planmeca Compact i dental unit should assure that it is
used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
A-8 Planmeca Compact i
Guidance and manufacturer's declaration - electromagnetic emissions
Compliance
Group 1
Class B
Class A
Complies
Electromagnetic environment – guidance
Planmeca Compact i dental unit uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
Planmeca Compact i dental unit is suitable for
use in all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
EMC INFORMATION