Chapter C - SERVICE MODE
n.99
The user can select either a short or a long instrument hose flushing. Use this service mode
to enable/ disable the HE99 message.
In case the long flushing is not performed prior to unit turn-off:
0 = HE99 is not displayed
1 = HE99 will be shown, but the unit can be used without performing the long flushing.
2 = HE99 will be shown and the long flushing must be performed before the unit can be used.
The factory default is 0.
n.100 Select dental unit type (n mode = 0, 1, 2, 3, 4, 5, 6, 7)
This service mode is used to select the dental unit type.
0 = Prostyle Compact unit without chair
1 = Prostyle Compact unit
2 = Prostyle Compact WE'S unit
3 = reserved for future use
4 = Planmeca Compact e
5 = Planmeca Compact i without chair
6 = Planmeca Compact i
7 = Planmeca Compact e with swivel position recognition
8 = Planmeca Compact i unit without chair, connected to Planmeca Chair
9 = Planmeca Simulation Unit
---- = unit type not selected
The factory default is 6.
After replacing the Main PCB the dental unit automatically enters to this service mode when
unit is switched on. The unit type must be selected before it can be used.
n.101 Interrupt waterline cleaning operation (n = 0, 1, 2, 3, 4, 5, 6)
This service mode is used to check the state of the waterline cleaning operation.
0 = Normal operation; the waterline cleaning operation off.
1 - 6 = The waterline cleaning operation on. Different numbers present different phases:
1 = The waterline cleaning operation started. Waiting for the instrument logic to notify the
operation start.
2 = The instrument logic has notified the operation start. The unit checks which doriot arms
are bent or removed. The unit is waiting for the container to be installed. Spreading the clean-
ing solution to the system.
3 = The cleaning solution has filled the instrument waterlines and the glass filler line. The unit
is waiting for the container to be removed and filled with water. After attaching the container to
the unit the flushing of the unit is being carried out.
4 = The rinsing cycle is ready. Waiting the container to be removed.
5 = Performing the final flush operation to clean the container properly. Going to the normal
operation state (number 0) after this.
6 = Some failure or error or user operation has occurred that might leave the unit dirty and the
unit must be flushed by pressing the flush button. Going to the normal operation state
(number 0) after this.
Whenever wanted, by pressing and holding down the S-key the waterline cleaning operation
can be interrupted. After interrupting the operation the text "turn off" will flash on the display.
Now turn off the unit.
The factory default is 0.
C-38 Planmeca Compact i
Instrument hose flushing operation: disable HE 99 message
SERVICE MODE DETAILED DESCRIPTION