Planmeca Compact I Technical Manual page 34

Table of Contents

Advertisement

Chapter C - SERVICE MODE
MODE
n.39
Set maximum fibre optic and Intracam bulb voltage for
selected hose type
n.40
Fine-adjust bulb voltage for selected polymerization light ---- = no valid light selected
n.41
Set fibre optic light turn-off delay for all instruments
n.42
Fine-adjust cooling air pressure for selected hose type
n.43
Set mode of foot controlled chip-blow
n.44
Fine-adjust drive air pressure for selected hose type
MODE
n 45
Show control panel type/AUX input for branch 1
n 46
Show control panel type/AUX input for branch 2
n 47
Show control panel key code for branch 1
n 48
Show control panel key code for branch 2
n 49
Display test
C-6 Planmeca Compact i
Table 3: Instrument related service modes
SERVICE MODE FUNCTION
Table 4: Control panel related service modes
SERVICE MODE FUNCTION
GENERAL ABOUT THE SERVICE MODE
DISPLAY/RANGE
---- = no valid hose selected
0 = factory default (nominal volt-
age)
< 0 = decrease max. voltage
> 0 = increase max. voltage
0 = factory default
> 0 = increase voltage
< 0 = decrease voltage
nn = seconds; factory default =
30 s.
---- = no valid hose selected
0 = factory default
< 0 = decrease cooling air pres-
sure
> 0 = increase cooling air pres-
sure
0 = as auto-chip is programmed
1 = always dry (factory default)
2 = no chip blow
---- = no valid hose selected
0 = factory default
< 0 = decrease drive pressure
> 0 = increase pressure
DISPLAY/RANGE
- = no panel found
0 = nurse's plus panel
1 = doctor's panel
2 = nurse's panel
3 = doctor's plus panel
- = no panel found
0 = nurse's plus panel
1 = doctor's panel
2 = nurse's panel
3 = doctor's plus panel
n.- = row, -.n = column
n.- = row, -.n = column
All LEDs turn on
page
C-26
C-26
C-26
C-26
C-26
C-26
page
C-27
C-27
C-27
C-27
C-27
Technical Manual