SERVICE MODE DETAILED DESCRIPTION
n. 137 Bowl rotation safety features
n.138 Simulation system type
n.139 Duration of long instrument hose flushing
n.140 Suction valve closing time adjustment
n 200 Boot information bits
n 201 Boot software checksum
n 202 Main software checksum
n.203 Processor watchdog test
n.204 Processor reset test
n.205 Lifter height calibration in millimetres (mm)
n.206 Backrest angle calibration in degrees (deg)
Technical Manual
In this mode the bowl rotation safety operation (when chair is driven upwards) is defined. If
the dental unit is not equipped with bowl, the bowl rotation is not included in the safety circuit.
0 = Bowl rotation is not included in the safety circuit
1 = Bowl rotation is included in the safety circuit (default)
Use this mode to select the Simulation system type.
0 = IDSS - Interactive Digital Simulation System
1 = ASW-1 - Advanced Simulation Workplace Single (Main Valves are in normal use)
2 = ASW-N - Advanced Simulation Workplace (Main Valves are not in use)
Use this mode to set the duration of the long instrument hose flushing. The adjustment range
is 2 - 9 minutes. Factory default is 2 minutes.
Use this mode to set the suction valve closing time. The adjustment range is 0 - 3 seconds.
Factory default is 0.
Reserved for factory use.
Reserved for factory use.
Reserved for factory use.
Restarts the dental unit with error E 75.1.
Restarts the dental unit without errors.
Measure from floor to the center of the backrest joint and set that value to this service mode.
Measure angle (or drive to 0 with spirit level) and set that angle here.
Chapter C - SERVICE MODE
Planmeca Compact i C-45