ELECTRICAL PARTS REPLACEMENT
5.2 Replacing fuses
WARNING
Fuse Panel
Code
F5
F6
F7
F8
F9
F10
Ratings for Mains Fuse Neutral (F11) and Mains Fuse Live (F12)
NOTE
Mains
Voltage
Setting
100V~
115V~
220-240V~
Technical Manual
Always replace fuses with the same type and rating.
Fuse Description
Operating light fuse
Separator 24V~ fuse
Water heater 24V~ fuse
Syringe heater 24V~ fuse
Electronics +24V Fuse A
Electronics +24V Fuse B
The mains fuse ratings depend on the selected voltage setting.
Fuse Rating & Type
Schurter T12.5A, slow blow,
Schurter T12.5A, slow blow,
T6.3A, slow blow
Chapter G - PARTS REPLACEMENT & REPAIR
Fuse Rating & Type
0034.3128
0034.3128
T6.3AL, slow blow
T8AL, slow blow
T6.3AL, slow blow
T6.3AL, slow blow
T4AL, slow blow
T4AL, slow blow
Planmeca Compact i G-21
Need help?
Do you have a question about the Compact I and is the answer not in the manual?
Questions and answers