Planmeca Compact I Technical Manual page 148

Table of Contents

Advertisement

Chapter F - ADJUSTMENTS
Bottle filling and emptying is repeated 3 times to clean the bottle
City Water
Flushing cycle:
Bottle flushing, filling
V1
o
V2
o
V3
c
V4
c
V5
c
Pump
running
Display
wait
Duration
Until L1 is reached
Bottle flushing, emptying
V1
c
V2
c
V3
o
V4
o
V5
c
Pump
not running
Display
15, when L2 reached 15 s count-
down timer
Duration
until L2 reached + 15 s
Bottle is filled
City Water
Filling bottle
V1
o
V2
o
V3
c
V4
o
V5
c
Pump
running
Display
wait
Duration
until L1 reached
F-38 Planmeca Compact i
Bottled Water
Flushing cycle:
Bottle is filled manually
He 65
Bottle flushing, emptying
V1
c
V2
c
V3
o
V4
o
V5
c
Pump
not running
Display
wait, 15 s countdown timer
Duration
until L2 reached + 15 s
Bottled Water
Bottle is filled manually
He 65
WMS OPERATION
Technical Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents