Planmeca Compact I Technical Manual page 146

Table of Contents

Advertisement

Chapter F - ADJUSTMENTS
WCS is started with chip blow key and the bottle has to be filled with cleaning agent. After that
the feeding cycle starts.
Cleaning agent is fed to instruments
City Water
SW1
off
Feeding cycle:
Instruments
V1
c
V2
c
V3
o
V4
c
V5
c
Pump
not running
Display
countdown timer
Duration
syringe 40 s, rightmost instrument 5
s, following instruments 3 s each
Cleaning agent is fed to cup fill line
City Water
Feeding cycle:
Cup fill
V1
c
V2
c
V3
o
V4
c
V5
o
Pump
not running
Display
countdown timer 4 s
Duration
4 s
F-36 Planmeca Compact i
Bottled Water
SW1
on
Similar to City water mode
Bottled Water
Similar to City water mode
WMS OPERATION