3 EMC INFORMATION

WARNING
Use of any accessories and cables other than those specified in Planmeca ProX
intraoral X-ray unit's documentation, with exception of cables sold by Planmeca as
replacement parts for internal components, may result in increased emission or
decreased immunity of the X-ray unit.
WARNING
Planmeca ProX intraoral X-ray unit should not be used adjacent to or stacked with
other equipment. If adjacent or stacked use is necessary, the Planmeca ProX
intraoral X-ray unit should be observed to verify normal operation in configuration
which it will be used.
Guidance and manufacturer's declaration - electromagnetic emissions
Planmeca ProX intraoral X-ray unit is intended for use in the electromagnetic environment specified
below. The customer or the user of the Planmeca ProX intraoral X-ray unit should assure that it is used
in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Technical manual
Compliance
Planmeca ProX intraoral X-ray unit uses RF energy
only for its internal function. Therefore, its RF
Group 1
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Planmeca ProX intraoral X-ray unit is suitable for
use in all establishments, including domestic
Class B
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Class A
Complies
Electromagnetic environment – guidance
3 EMC INFORMATION