SERVICE MODE DETAILED DESCRIPTION

2 SERVICE MODE DETAILED DESCRIPTION

2.1 MAIN control system related service modes

n.0
Technical Manual
The following pages contain detailed information about the service modes, giving useful infor-
mation on how to use and interpret the information available. Reading this chapter gives a
good understanding of the unit's control system. Used abbreviations are explained in the
beginning of this manual.
Underlined number or values in the heading texts are factory default values.
Disable programming and limit micromotor speed
(For educational establishment use.) In this mode the programming of the instrument, unit
and/or chair settings can be disabled, and the micromotor power can be limited to 50% of the
maximum speed.
The following settings are possible. The factory default is 0 (zero).
instrument
settings
disabled
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
unit settings
disabled
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
Chapter C - SERVICE MODE
chair settings
micromotor
disabled
speed limited
x
x
x
x
x
x
x
x
Planmeca Compact i C-17
x
x
x
x
x
x
x
x

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Compact I and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Table of Contents

Save PDF