WMS OPERATION
Cup fill+bowl rinse lines and instruments are flushed and hereafter bottle is filled. This is
repeated 4 times.
Countdown timer starts from 4 min. (4 x 1 min.). Timer is stopped during bottle filling.
City Water
Flushing cycle:
Cup fill + bowl rinse flushing
V1
o
V2
o
V3
o
V4
c
V5
o
Pump
running
Display
timer
Duration
10 s (at same time bowl rinse, V6
open 3 s)
Flushing cycle:
Instrument flushing (each instrument 10 s by
turns, syringe all time)
V1
o
V2
o
V3
o
V4
c
V5
c
Pump
running
Display
timer
Duration
50 s
Flushing cycle:
Bottle filling
V1
o
V2
o
V3
c
V4
c
V5
c
Pump
running
Display
wait
Duration
Until L1 reached
When cycle is completed
Display: He 36
Technical Manual
Bottled Water
Similar to city water mode
Similar to City water mode
Bottle has to be filled manually
He 65
Chapter F - ADJUSTMENTS