Chapter C - SERVICE MODE
MODE
n.84
Set type of extra assistant instrument
Table 10: MCB CPU direct I/O related service modes (for factory use)
MODE
n 85
Show IC36 output signals
n 86
Show IC37 output signals
n 87
Show IC29 driver signals
n 88
Show IC31 driver signals
n 89
Show IC14 input signals
n 90
Show IC16 input signals
n 91
Show port A outputs (IC4)
n 92
Show port A inputs (IC4)
MODE
n.93
Adjustment of the duration of the spray water impulse
n.94
Set syringe fibre optic light voltage
n.95
Set spray water impulse pressure
n.96
RPM display / Enable low speed function for brush micro-
motors (Optional Plus package)
n.97
Set suction system cleaning time
C-10 Planmeca Compact i
Table 9: Separator/suction system related service modes
SERVICE MODE FUNCTION
SERVICE MODE FUNCTION
Table 11: Other service modes
SERVICE MODE FUNCTION
GENERAL ABOUT THE SERVICE MODE
DISPLAY/RANGE
0 = no extra assistant instrument
1 = assistant's syringe
2 = water operated saliva ejector
3 = assistant's syringe and water
operated saliva ejector
4 = PM Starlight LED polymeriza-
tion light
5 = assistant's syringe and PM
Starlight
6 = water operated saliva ejector
and PM Starlight
7 = assistant's syringe, water
operated saliva ejector and PM
Starlight
DISPLAY/RANGE
¦¦¦¦¦¦¦¦
¦¦¦¦¦¦¦¦
¦¦¦¦¦¦¦¦
¦¦¦¦¦¦¦¦
¦¦¦¦¦¦¦¦
¦¦¦¦¦¦¦¦
¦¦¦¦¦
¦¦¦
DISPLAY/RANGE
0 - 20, default = 1
0 - 4 V, default = 3.7 V
0 - 30, default = 15
0 = RPM disabled / low speed
function disabled
1 = RPM disabled / low speed
function enabled
2 = RPM enabled / low speed
function disabled
3 = RPM enabled / low speed
function enabled
0 sec. - 20 min, default = 2 min