GENERAL & TECHNICAL DATA
1.6

EMC information

WARNING
Use of any accessories and cables other than those specified in
Planmeca Intra X-ray unit's documentation, with exception of
cables sold by Planmeca as replacement parts for internal
components, may result in increased emission or decreased
immunity of the X-ray unit.
WARNING
Planmeca Intra X-ray unit should not be used adjacent to or
stacked with other equipment. If adjacent or stacked use is
necessary, the Planmeca Intra X-ray unit should be observed to
verify normal operation in configuration which it will be used.
Guidance and manufacturer's declaration - electromagnetic emissions
Planmeca Intra X-ray unit is intended for use in the electromagnetic environment specified
below. The customer or the user of the Planmeca Intra X-ray unit should assure that it is used in
such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
8
Planmeca Intra X-ray unit
Compliance
Electromagnetic environment – guidance
Planmeca Intra X-ray unit uses RF energy only
for its internal function. Therefore, its RF
Group 1
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
Planmeca Intra X-ray unit is suitable for use in
all
Class B
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Class A
Complies
establishments,
including
domestic