WMS OPERATION
Cleaning agent is fed to internal water lines
City Water
Feeding cycle:
Internal water lines
V1
c
V2
o
V3
o
V4
o
V5
c
Pump
running
Display
countdown timer 30 s
Duration
30 s
Excess cleaning agent is removed from bottle
City Water
Feeding cycle:
Emptying bottle
V1
c
V2
c
V3
o
V4
o
V5
c
Pump
not running
Display
wait, when L2 reached countdown
timer 15 s
Duration
until L2 reached + 15 s
When feeding cycle is completed, Display: He 66
Flushing cycle is started
City Water
After 8 hours if unit let on or when unit restarted
on
Technical Manual
Bottled Water
Similar to City water mode
Bottled Water
Similar to City water mode
Bottled Water
When unit turned on