Planmeca Compact I Technical Manual page 86

Table of Contents

Advertisement

Chapter D - HELP & ERROR MESSAGES
Error
E27
reserved
INSTRUMENT MULTIPLEXER RELATED
E28.n
Error in instrument hose identification system in IMUX.
E29
Voltage error in fibre light power supply in IMUX.
E30
reserved
E31
reserved
E32.n
Error in serial communication between MCB and IMUX.
E33.n
IMUX output short circuit (valve or wire short circuit).
E34.n
IMUX output open-circuit (valve or wire open-circuit).
E35.n
IMUX pressure sensor offset error (signal outside range).
E36.n
IMUX pressure sensor error (signal voltage too high).
E37
Pressure sensor cable not properly attached to the IMUX.
E38
reserved
E39
Drive (cooling) air valve control servo cannot maintain pressure to instrument.
E40
reserved
E41
reserved
E42
Air coolant (spray air) valve control servo cannot maintain pressure to instru-
ment.
E43
reserved
E44
reserved
E45
Water coolant (spray water) valve control servo cannot maintain pressure to
instrument.
E46
reserved
SUCTION TUBE HOLDER RELATED
E47.n
Suction tube holder is of new type. Update the MCB software.
E48.n
Number of attached suction tube holders is wrong.
E49.n
Suction tube holder cable or electronics is faulty.
E50
reserved
FOOT CONTROL RELATED
E51.n
Cable from the foot control is loose
E52
Foot control is of a new type. Update software on MCB.
E53
Foot control serial communication (data) failure.
E54.n
Foot control is not properly calibrated.
E55
reserved
D-8 Planmeca Compact i
Explanation
ERROR MESSAGES
Details on:
page D-14
page D-14
page D-14
page D-15
page D-15
page D-15
page D-16
page D-16
page D-16
page D-16
page D-17
page D-17
page D-17
page D-17
page D-18
page D-18
page D-18
page D-18
Technical Manual