Connecting The Stream-Switching Valve (Corona Veo Rs Only) - Thermo Scientific Corona Veo Operating Manual

Charged aerosol detectors
Table of Contents

Advertisement

5
Installation
5.6.6

Connecting the Stream-Switching Valve (Corona Veo RS only)

Parts required
Page 96
The Corona Veo RS detector is equipped with a 6-port 2-position stream-
switching valve (SSV). The valve can be used to direct flow to the
nebulizer or to external devices or to a waste reservoir.
To connect the stream-switching valve, connect ports no. 2, 3, and 4 of
the valve. The ports of the stream-switching valve are assigned as
follows:
Figure 15: Ports on the stream-switching valve
No.
Ports
2
Divert/Waste
Outlet to direct flow away from the detector
3
Inlet
Inlet for capillary from column
4
Nebulizer
Outlet to direct flow to the nebulizer inlet
1, 5, 6
Not used
From the detector ship kit:
RheFlex fittings
Inlet capillary, PEEK, 0.005" x 1/16" (I.D. x O.D.), red
Interconnecting capillary, Viper, stainless steel,
0.13 x 100 mm (I.D. x L)
Waste capillary, PEEK, 0.015" x 1/16" (I.D. x O.D.), gray
Charged Aerosol Detectors Corona Veo (RS) Operating Manual