Disinfection; Rinsing; Uv Disinfection After Cleaning - Thermo Scientific 1500 A2 Series Operating Manual

Class ii, type a2 biological safety cabinet
Table of Contents

Advertisement

Disinfection

UV Disinfection After
Cleaning
Thermo Scientific
1. Remove all samples from the chamber and store them properly.
2. Remove accessories from the cabinet and disinfect them using the
disinfection procedure recommended by the manufacturer of the
accessory.
3. The work tray(s) and stainless steel components can be removed from
the chamber and disinfected separately.
4. Clean all chamber surfaces with disinfectant.
5. Do not remove the optional UV lights from the sockets. Clean
thoroughly with a damp cloth. Do not saturate. Be sure to dry
completely.

Rinsing

1. Rinse all surfaces twice with clean distilled water.
2. Discard any liquid from the drain pan. Rinse pan.
3. Dry all chamber surfaces completely.
UV disinfection can be performed by using the optional factory installed
UV light.
To start the UV disinfection procedure,
1. Close the front window, the air system operates in the reduced mode.
2. Press the UV key on the control panel.
To interrupt or cancel the UV disinfection procedure, just press the UV
key (the display shows the operating hours) and slide the window up.
Section 5
Cleaning and Decontamination
1500 Series A2
5-3