Illegal Temperature Value from Sensor at Air Inlet

Error ID: 1072
This temperature sensor delivered a value outside the allowed range. The
parameter of this event equals the measured temperature in 1/100 centigrade.
As a result the temperature control is switched off.
Probable cause
1
2

Wavelength Recalibration Lost

Error ID: 1037
The calibration information needed for your detector to operate correctly has
been lost.
During calibration of the detector the calibration values are stored in ROM. If
no data is available in the spectrometer ROM, the error message is generated.
Probable cause
1
2
Agilent 1260 Infinity DAD and MWD User Manual
The temperature sensor is defect.
Detector is exposed to illegal ambient
conditions.
The detector is new.
The detector has been repaired.
Error Information
Detector Error Messages
Suggested actions
• Replace the cable to the main board.
• Please contact your Agilent service
representative.
Verify that the ambient conditions are within
the allowed range.
Suggested actions
Recalibrate the detector.
Please contact your Agilent service
representative.