Probable Causes
Suggested Actions
Troubleshooting and Test Functions

Needle Up Failed

Needle Up Failed
The needle arm failed to move successfully from the seat or out of the vial to
the upper position.
The upper position of the needle arm is monitored by a position sensor on the
sampling unit flex board. The sensor detects the successful completion of the
needle movement to the upper position. If the needle fails to reach the end
point, or if the sensor fails to recognize the needle arm movement, the error
message is generated.
• Defective or dirty position sensor.
• Defective motor.
• Sticking spindle assembly.
• Defective ASM board.
❏ Exchange the sampling unit flex board.
❏ Exchange the needle drive motor.
❏ Exchange the spindle assembly or sampling unit assembly.
❏ Exchange ASM board.