Error Messages

AI Door Error

Possible Causes
The Automation Interface door is blocked
A door sensor has failed

Plate Error

Possible Causes
The plate does not move correctly
The transport belt is dirty or wet
The transport belt is damaged
The transport belt sensor has failed
The plate is stuck in the gripper of the Well-plate
Handler

AI Door Sensor Error

Possible Causes
The Automation Interface door was opened
manually
An Automation Interface door sensor is defective Replace the Automation Interface
Agilent 1200 Series Sample Capacity Extension User Manual
Troubleshooting and Error Handling
Actions
Remove any blockages
Replace the Automation Interface
Actions
Remove the plate
Clean the transport belt
Replace the Automation Interface
Replace the Automation Interface
Use the Automation Interface test function to
open the gripper of the Well-plate Handler
Actions
Do not open the Automation Interface door
manually, always use the Automation Interface
test function to open the door