The Agilent 1260 Infinity Binary LC - Product Description

Features of the Agilent 1260 Infinity Binary LC

Features of the Agilent 1260 Infinity Binary LC
The design concept of the 1260 Infinity Binary LC is to provide a liquid
chromatograph offering ultra fast and high resolution separation capability
and yet which retains full functionality for standard HPLC applications.
Therefore, it provides full backwards compatibility for your established
HPLC and UHPLC methods. The use of sub- two micron (STM) particles
means that for high flowrates or long columns additional pressure is
required to drive the mobile phase through the column. The flowpath of
the 1260 Infinity Binary LC is optimized to produce minimal backpressure,
and ZORBAX RRHT columns have an engineered particle size distribution
that produces significantly less backpressure than other STM columns.
The design features and benefits of the Agilent 1260 Infinity Binary LC
are:
• The configurable delay volume down to 120 µL in the 1260 Infinity
• The standard delay volume configuration of the 1260 Infinity Binary
• The next generation flow- through design of the 1260 Infinity High
• High temperature, up to 100 °C on certain columns, allows more
• In the 1290 Infinity Thermostatted Column Compartment , different
8
Binary Pump combined with a flow range from 0.05 up to 5 mL/min at
pressures up to 600 bar provides universal applicability from narrow
bore (2.1 mm ID) to standard bore (4.6 mm ID) columns, matching the
needs for both LC and LC/MS.
Pump allows you to run not only UHPLC but also conventional HPLC
methods without compromising performance or changing
chromatographic patterns.
Performance Autosampler achieves highest precision for a wide range of
injection volumes (from 0.1 up to 100 µL) without changing sample
loops. It is designed for high sample throughput, low carryover, and fast
injection cycles.
selectivity flexibility and reduces solvent viscosity to allow even faster
separation.
heater (1.6 µL) and cooling elements (1.5 µL) for low extra- column
volume can be installed. The temperature is adjustable from 10 °C
below ambient up to 100 °C.
1260 Infinity Binary LC - System User Guide