Run
This button initiates the selected diagnostic tests. During the diagnostic test run, the screen
appears as in
Figure
61. At completion of the test, the
Diagnostic Test Report screen
opens.
Close
This button returns you to the Instrument Diagnostic screen.
Diagnostic Test Report screen
Purpose: To view the results of instrument diagnostic tests. See instructions in
"View reports for
diagnostic tests and instrument health checks"
on page 40.
To open: From the Home screen, press Settings. On the Settings screen press Self Diagnostic,
then press Browse Report. to open the Diagnostic Report Explorer screen. Then, select a report
and press View.
Note that this screen opens automatically after completion of diagnostic testing.
Figure 62 Diagnostic Test Report screen
Report table
The table in the center of the screen lists the diagnostic tests that were performed and the result
of the test (Passed, Failed, or Skipped).
When viewing the diagnostic test report for a test that just completed, if any of the items failed,
you will see an error icon near the bottom of the screen, like the one shown below. To view more
information about the failed test items, press directly on the icon at the bottom of the screen.
Close
This button returns you to the Diagnostic Report Explorer screen.
Magnis and MagnisDx NGS Prep System User Guide
82