Running The Test - Agilent Technologies 1260 Infinity VL User Manual

Quaternary pump
Table of Contents

Advertisement

Running the Test

Parts required
p/n
01080-83202
Damage to pressure sensitive parts
C A U T I O N
Even columns that are suitable for high pressures are sensitive to pressure drops that
occur during this test.
Running the test from the Agilent Lab Advisor
1 Select the System Pressure Test from the Test Selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by slowly opening the purge valve when the test has
N O T E
finished.
"Evaluating the Results"
interpretation of the System Pressure Test results. For detailed instructions
refer to the Agilent Lab Advisor software.
1260 Infinity QuatPump VL User Manual
Description
Blank nut
Do not include any pressure sensitive parts to the flow path and choose a maximum
pressure that is compatible to your system. For example, do not include columns, a
standard pressure flow cell (up to 20 bar) or a 400 bar autosampler to a 600 bar
pressure test.
on page 100 describes the evaluation and
Test Functions and Calibration
System Pressure Test
8
99