Serial Bus Decoding & Triggering, Search & Navigation, and Segmented Acquisition Labs
Lab #12: Decoding, Triggering, and Searching on RS232/UART Serial Bus
Signals
In this lab you will learn how to set up the scope to decode and trigger on
transmit and receive RS232/UART serial bus traffic. In addition, you will
learn how to use the scope's automatic RS232/UART Search & Navigation
capability, as well as Segmented Memory acquisition. This lab does not
provide a tutorial on the RS232/UART protocols and signaling. To complete
this lab, your scope must be licensed with the RS232/UART trigger and
decode option (Option COMP). You can verify the installed options on your
oscilloscope at [Help] > About Oscilloscope.
1
2
3
4
5
6
7
8
9
10
11
12
2000 X-Series Oscilloscopes Advanced Training Guide
Connect the channel- 1 probe to the Demo 1 terminal and ground.
Connect the channel- 2 probe to the Demo 2 terminal and ground.
Press [Default Setup] on the scope's front panel.
Press [Help]; then press the Training Signals softkey.
Using the Entry knob, select the RS232/UART signal; then press the
Output softkey to turn it on.
Set channel- 1 to 1.00 V/div.
Set channel- 1's offset/position to -1 V.
Press the [2] front panel key to turn on channel- 2.
Set channel- 2 to 1.00 V/div.
Set channel- 2's offset/position to +2.3 V.
Push the trigger level knob to automatically set the trigger level to
approximately 50%.