Lab #19: Decoding, Triggering, and Searching on LIN Serial Bus Signals

In this lab you will learn how to set up the scope to decode and trigger on
a LIN serial bus signal. In addition, you will learn how to use the scope's
automatic LIN Search & Navigation capability, as well as Segmented
Memory acquisition. This lab does not provide a tutorial on the LIN
protocol and signaling, which is primarily used in automotive applications.
To complete this lab, your scope must be licensed with the CAN/LIN
trigger and decode option (Option AUTO). You can verify the installed
options on your oscilloscope at [Help] > About Oscilloscope.
1
2
3
4
5
6
7
8
Figure 73 Capturing a LIN frame using the scope's Edge triggering mode.
3000 X-Series Oscilloscopes Advanced Training Guide
Connect the channel- 1 probe to the Demo 1 terminal and ground.
Press [Default Setup] on the scope's front panel.
Press [Help]; then press the Training Signals softkey.
Using the Entry knob, select the LIN signal; then press the Output
softkey to turn it on.
Set channel- 1 to 500.0 mV/div.
Set channel- 1's offset/position to +1.4 V.
Push the trigger level knob to automatically set the trigger level to
approximately 50%.
Set the scope's timebase to 1.00 ms/div.