2
Testing Performance
To verify bandwidth
Table 5
Equipment Required to Verify Bandwidth
Equipment
Signal Generator
Power Meter/Sensor
Power Splitter
Cable
Adapter
28
This test checks the bandwidth of the oscilloscope. In this test
you will use a signal generator and a power meter.
500 MHz Models
Test limits at 2 mV/div to 5 V/div
• All channels (±3 dB)
•
dc to 500 MHz
300 MHz Models
Test limits at 2 mV/div to 5 V/div
• All channels (±3 dB)
•
dc to 300 MHz
100 MHz Models
Test limits at 1 mV/div to 5 V/div
• All channels (±3 dB)
•
dc to 100 MHz
Critical Specifications
100 kHz - 500 MHz at 200 mVrms
1 MHz - 1 GHz ±3% accuracy
Outputs differ by ≤ 0.15 dB
Type N (m) 24 inch
Type N (m) to BNC (m)
Recommended
Model/Part
Agilent N5181A, Agilent
E4400B, or Agilent 8648A
Agilent E4418B/8482A
Agilent 11667A
Agilent 11500B
Agilent 1250-0082
5000 Series Oscilloscopes Service Guide
Qty
1
1
1
1
1