Table 8. Equipment Required To Verify Bandwidth - Agilent Technologies 6000 Series Service Manual

Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

2
Testing Performance
42
Table 8
Equipment Required to Verify Bandwidth
Equipment
Critical Specifications
Signal Generator
100 kHz - 1 GHz at 200 mVrms
Power Meter/Sensor
1 MHz - 1 GHz ±3% accuracy
Power Splitter
outputs differ by < 0.15 dB
Cable
Type N (m) 24 inch
Adapter
Type N (m) to BNC (m)
Feedthrough
50Ω BNC (f) to BNC (m)
1 Connect the equipment (see
a Use the N cable to connect the signal generator to the
input of the power splitter input.
b Connect the power sensor to one output of the power
splitter.
c For 300 MHz, 500 MHz, and 1 GHz oscilloscope models use
an N-to-BNC adapter to connect the other splitter output
to the channel 1 input.
d For 100 MHz oscilloscopes, use an N-to-BNC adapter and
50Ω feedthrough termination to connect the other splitter
output to the channel 1 input on the oscilloscope.
Recommended
Model/Part
Agilent N5181A, or
E4400B, or 8648A
Agilent E4418B/8482A
Agilent 11667A
Agilent 11500B
Agilent 1250-0082
Agilent 0960-0301
Figure
5).
6000 Series Oscilloscopes Service Guide