4
Programming Command Guidelines
Overview of the GP-IB Interface
Table 4-2 GP-IB Configuration
Item
GP-IB Address:
EOI
Terminator:
4-6
Settings for the GP-IB Interface
Refer to the documentation on your controller, programming
language, and I/O card to configure the GP-IB interface.
The GP-IB configuration is not programmable but manually set.
The preset settings and other choices are as follows:
Preset
2
Off
CR+LF
The terminator outputted from the Tester can be changed as
specified in
Table
The terminator inputted to the Tester is LF (fixed).
If EOI is set to
On
Refer to the following procedure to set the GP-IB configuration.
Step 1
On the [Initial] screen, press the Config softkey to obtain the
[Configuration] screen.
Allowable Range
1 to 15
On, Off
CR+LF, CR, LF
4-2.
, the terminator is not transmitted.
N9360A Multi UE Tester W-CDMA Programming Manual