Programming Command Reference
5-14
<real 5>,<real 6>
CONFigure:LOSS:RFOUt?
This command specifies the losses between a mobile phone and
the Tester caused by the RF Input path for each band,<real 1>
Band 1 and <real 6> Band 6, respectively. <real 2>, <real
3>, <real 4> and <real 5> is unused (Set to
allowable range is from
these are set to
dB.
0.0
This query command returns the insertion loss setting with real
value. Each value is separated by a comma. The range of these
values is same as that of the command parameter.
Beeper Control
CONFigure:BEEPer ON|OFF|1|0
CONFigure:BEEPer?
This command sets the beeper function to ON (1) or OFF (0). If
this is set to ON, there is beep for each step of operation. If this
is set to OFF, beep is suppressed except for some errors and
warnings. At preset, this is set to ON.
This query command returns the beep setting with integer value
of
or
.
1
0
10MHz Reference Oscillator
CONFigure:ROSCillator?
This query command returns the reference signal currently
used, which is either INT for the internal reference or EXT for
the external reference.
Date
CONFigure:DATE <int YYYY>,<int MM>,<int DD>
CONFigure:DATE?
This command sets the default of the calendar in the Tester.
<int YYYY> is year within the range from
MM> is month within the range from
day within the range from
N9360A Multi UE Tester W-CDMA Programming Manual
to
dB in
0.0
99.9
0.1
1990
to
and <int DD> is
1
12
to
.
1
31
). The
0.0
steps. At preset,
to
, <int
2037