Advertisement

5
Programming Command Reference
5-26
This command defines whether or not to include the Local
Update in a test flow. Setting
On Location Update,
and
skips Location Update test. At preset, this is set to
OFF
in sequence 1 and
LU2
MS Call
CONFigure:SEQuence:SIGNaling:MSCAll ON|OFF|1|0
CONFigure:SEQuence:SIGNaling:MSCAll?
This command sets whether or not to include the MS Call in a
test flow. The parameter is
At preset, this is set to
This query command returns the setting of MS Call with integer
value of
or
. The meaning of the return value is same as that
1
0
of the command parameter.
Talk Test (MS Call)
CONFigure:SEQuence:SIGNaling:TALK ON|OFF|1|0
CONFigure:SEQuence:SIGNaling:TALK?
This command sets whether or not to include the Talk step with
an AMR connection (MS Call) in a test flow. The parameter is
(1) for Run or
(0) for — (skip). At preset, this is set to
OFF
This query command returns the setting of Talk test with
integer value of
or
1
0
as that of the command parameter.
BS Call (AMR)
CONFigure:SEQuence:SIGNaling:BSCAll:<<
AMR
ON|OFF|1|0
CONFigure:SEQuence:SIGNaling:BSCAll:AMR?
This command sets whether or not to include the BS Call (AMR)
in a test flow. The parameter is
(skip). At preset, this is set to
N9360A Multi UE Tester W-CDMA Programming Manual
executes Mobile Phone Power
Lu1
executes BS Change Location Update
Lu2
in sequence 2.
(1) for Run or
ON
.
1
. The meaning of the return value is same
(1) for Run or
ON
.
1
LU1
(0) for — (skip).
OFF
ON
.
1
(0) for —
OFF