Command Reference for RF Test Result

N9360A Multi UE Tester W-CDMA Programming Manual
bit 8(256): Frequency Error
bit 9(512): EVM
bit 10(1024):Sensitivity/BER
bit 11(2048): Abort
Each bit is set to
if the test is pass or skipped, or each bit is set
0
to
if the test is fail.
1
If there are no test results, all bits are set to
returned.
The value in the parentheses is decimal value for each bit. Refer
to
Setting and Querying the Registers
The following query commands are used to return a set of the
RF test result.
Open Loop TX Power Measurement
TESTs:AUTO:RESults:OPENpower1?
This query command returns the measurement result values,
<int 1>, <int 2> and <real 3>, for RFCH 1.
• <int 1> (1 digit) is
measurement range, or
range or the Open Loop TX Power is not measured.
• <int 2> (1 digit) is the test result for the pass (0) or fail (1).
• <real 3> (up to 5 letters) is the measured data ranging
from
to
-99.9
+99.9
range or the Open Loop TX Power is not measured, the value
is returned.
-99.9
The Open Loop TX Power measurement can be executed only on
RFCH1.
Inner Loop Power Measurement (Down MIN)
TESTs:AUTO:RESults:INNErpower1:DownMIN?
TESTs:AUTO:RESults:INNErpower2:DownMIN?
TESTs:AUTO:RESults:INNErpower3:DownMIN?
Programming Command Reference
1
on
page
if measurement result is within the
0
if measurement result is out of
1
in dBm. If measurement result is out of
5
, the value
is
4095